    zenith_flex_aaa  (excerpted from IFU)

    MRI Information
    Non-clinical testing has demonstrated that the Zenith AAA Endovascular Graft is MR Conditional. A patient with this endovascular graft can be scanned safely immediately after placement under the following conditions:

    Static Magnetic Field

    • Static magnetic field of 3.0 Tesla or less.
    • Highest spatial magnetic gradient field of 720 gauss/cm

    Non-clinical evaluation was conducted in a 3 Tesla MR system (General Electric Excite) with a maximum spatial magnetic gradient field of 720 gauss/cm as measured with a gaussmeter in the position of the static magnetic field pertinent to the patient (i.e., outside of scanner covering, accessible to a patient or individual).

    MRI-Related Heating

    1.5 Tesla Systems:

    • Static magnetic field of 1.5 Tesla
    • Maximum whole-body-averaged specific absorption rate (SAR) of 2 W/kg for 15 minutes of scanning (i.e., per scanning sequence)

    In non-clinical testing, the Zenith AAA Endovascular Graft produced a temperature rise of less than or equal to 1.4°C at a MR system reported whole-body-averaged specific absorption rate (SAR) of 2.8 W/kg for 15 minutes of MR scanning in a 1.5 Tesla Magnetom, Siemens Medical Magnetom, Numaris/4 Software, Version Syngo MR 2002B DHHS MR Scanner. The maximum whole-body-averaged specific absorption rate (SAR) was 2.8 W/kg, which corresponds to a calorimetry measured value of 1.5 W/kg.

    3.0 Tesla Systems:

    • Static magnetic field of 3.0 Tesla
    • Maximum whole-body-averaged specific absorption rate (SAR) of 2 W/kg for 15 minutes of scanning (i.e., per scanning sequence)

    In non-clinical testing, the Zenith AAA Endovascular Graft produced a temperature rise of less than or equal to 1.9°C at a MR system reported whole-body-averaged specific absorption rate (SAR) of 3.0 W/kg for 15 minutes of MR scanning in a 3.0 Tesla Excite, GE Electric Healthcare, G3.0-052B Software, MR Scanner. The maximum whole-body-averaged specific absorption rate (SAR) was 3.0 W/kg, which corresponds to a calorimetry measured value of 2.8 W/kg.

    Image Artifact

    The image artifact extends throughout the anatomical region containing the device, obscuring the view of immediately adjacent anatomical structures within approximately 20 cm of the device, as well as the entire device and its lumen, when scanned in nonclinical testing using the sequence: Fast spin echo, in a 3.0 Tesla, Excite, GE Electric Healthcare, with G3.0-052B Software, MR system with body radiofrequency coil.

    For all scanners, the image artifact dissipates as the distance from the device to the area of interest increases. MR scans of the head and neck and lower extremities may be obtained without image artifact. Image artifact may be present in scans of the abdominal region and upper extremities, depending on distance from the device to the area of interest.

    Clinical information is available for seventeen patients who received MRI scans after stent-graft implantation. There have been no reported adverse events or device problems in any of these patients as a result of having received an MRI. Additionally, there have been well over 100,000 Zenith AAA Endovascular Grafts implanted worldwide, in which there have been no reported adverse events or device problems as a result of MRI.

    Cook recommends that the patient register the MR conditions disclosed in this IFU with the MedicAlert Foundation. The MedicAlert Foundation can be contacted in the following manners:

    Mail: MedicAlert Foundation International
    2323 Colorado Avenue
    Turlock, CA 95382
    Phone: 888.633.4298 (toll free)
    209.668.3333 from outside the U.S.
    Fax: 209.669.2450
    Web: www.medicalert.org
  • Cook Celect™ Vena Cava Filter

    Nonclinical testing has demonstrated that the Cook Celect Vena Cava Filter is MR Conditional. It can be scanned safely immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient field of 330 gauss/cm or less
    • Maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 15 minutes of scanning

    In nonclinical testing, the Cook Celect Vena Cava Filter produced a temperature rise of less than 0.2°C at a maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 15 minutes of MR scanning in a 3.0 tesla MR system using a transmit/receive body coil MR scanner. MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the Cook Celect Vena Cava Filter. Therefore, it may be necessary to optimize MR imaging parameters for the presence of this metallic implant.

  • Cook Medical adheres to the F2503-08 standard of MR (magnetic resonance) terminology adopted by the American Society for Testing and Materials (ASTM). This practice of marking medical devices came about in an effort to clarify the terminology, provide a uniform system for marking and, more important, because misuse of old terminology could result in serious injuries for patients and other individuals. This new system references the MR environment where testing took place, which includes field strength, spatial gradients, dB/dt (time rate of change of the magnetic field), radio frequency (RF) fields and specific absorption rate (SAR).

    Taking these conditions into consideration, three new categories of MR marking exist: MR Safe, MR Conditional, and MR Unsafe. In this system, “safe” and “unsafe” are the two extremes. “Conditional” is in the middle and characterizes the behavior of a medical device under specific conditions in the MR environment as determined by testing. In addition to the new terms, the ASTM International document introduced corresponding icons consistent with international standards for colors and shapes of safety signs. The icons are intended for use on items that may be brought into or near the MR environment as well as in product labeling. The new terminology and icon structure is intended to help clarify matters related to biomedical implants and devices to ensure the safe use of MRI technology.

    MR Safe logo

    MR Safe—An item that poses no known hazards in all MR environments. MR Safe items include nonconducting, nonmetallic, nonmagnetic items such as plastic, silicone and glass. An item may be determined to be MR Safe on the basis of scientific rationale rather than test data.

    MR Conditional logo

    MR Conditional—An item that has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use. Field conditions should be enumerated on the product, its packaging or in the enclosed literature. Additional conditions, including specific configurations of the item, may be required.

    MR Unsafe logo

    MR Unsafe—An item that is known to pose hazards in all MR environments. MR Unsafe items include magnetic items such as a pair of ferromagnetic scissors.

     

    Products Material Order Number
    (Prefix)
    MR
    Safe
    MR
    Conditional
    MR
    Unsafe
    AAA Endovascular Graft
         Zenith Flex®
            Visit product page »
    Stainless Steel TFFB, TFB, TFLE, ESBE, ESC, ESLE, ZIP, ESP   X  
         Zenith Renu®
            Visit product page »
    Stainless Steel AX1, RX1   X  
    Biopsy Needle
         Chiba Biopsy Needles
            Visit product page »
    Inconel® IDCHN   X  
    Embolization Coils
         Stainless Steel
            Visit product page »
    Stainless Steel MWCE     X
         Nester®
            Visit product page »
    Platinum MWCE-NESTER   X  
         MicroNester®
         Microcoils

            Visit product page »
    Platinum MWCE-NESTER   X  
         Tornado®
            Visit product page »
    Platinum MWCE-TORNADO   X  
         Hilal Microcoils
            Visit product page »
    Platinum MWCE-HILAL   X  
         MReye®
            Visit product page »
    Inconel IMWCE   X  
         MReye Flipper®
            Visit product page »
    Inconel IFMWCE   X  
    Reconstructive Surgery (Doppler Monitoring)
         Cook-Swartz Doppler
         Flow Probe

            Visit product page »
    Misc. Materials DP     X
    Stents
         Zilver 518® RX Vascular
            Visit product page »
    Nitinol ZIVX5   X  
         Zilver 518® Vascular
            Visit product page »
    Nitinol ZIV5   X  
         Zilver 518® Biliary
            Visit product page »
    Nitinol ZIB5   X  
         Zilver 635® Vascular
            Visit product page »
    Nitinol ZIV6   X  
         Zilver 635® Biliary
            Visit product page »
    Nitinol ZIB6   X  
         Zilver Flex 35 Biliary
            Visit product page »
    Nitinol ZIB6-35   X  
         Formula 418
            Visit product page »
    Stainless Steel FORB4   X  
    TAA Endovascular Graft
         Zenith® TX2®
            Visit product page »
    Stainless Steel ZTEG, ESBE   X  
    Vena Cava Filters
         Günther Tulip®
            Visit product page »
    Conichrome IGTCFS   X  
         Cook Celect®
            Visit product page »
    Conichrome IGTCFS-CELECT   X  
        Gianturco-
         Roehm Bird's Nest
    ®
            Visit product page »
    Stainless Steel BNF X    
    Vital-Port® Systems
         Dual Chamber
         Titanium–Petite

            Visit product page »
    Titanium IP   X  
         Dual Chamber
         Titanium–Standard

            Visit product page »
    Titanium IP   X  
         MRI–Petite
            Visit product page »
    Titanium IP   X  
         MRI–Standard
            Visit product page »
    Titanium IP   X  
         Titanium–Mini
            Visit product page »
    Titanium IP   X  
         Titanium–Petite
            Visit product page »
    Titanium IP   X  
         Titanium–Standard
            Visit product page »
    Titanium IP   X  
         Titanium for Peripheral
         Placement–Mini

            Visit product page »
    Titanium IP   X  

    Conichrome is a trademark of Carpenter Technologies.
    Inconel is a registered trademark of Huntington Alloys Corporation.
  • Zilver® Self-Expanding Stent

    Nonclinical testing has demonstrated that the Zilver Self-Expanding Stent is MR Conditional. It can be scanned safely under the following conditions:

    • Static magnetic field of 3 tesla or less
    • Spatial gradient field of 720 gauss/cm or less
    • Whole-body averaged specific absorption rate (SAR) of 1.5 W/kg (for a single stent at 1.5 tesla) and 3 W/kg (for a single stent at 3 tesla and a pair of overlapping stents at 1.5 and 3 tesla) for 20 minutes (for a single stent at 1.5 tesla) and 15 minutes of scanning (for a single stent at 3 tesla and a pair of overlapping stents at 1.5 and 3 tesla), respectively.

    In nonclinical testing, the Zilver Self-Expanding Stent produced maximum temperature rises of 0.1, 3.8, 0.8, and 0.1 degrees C (for a single stent at 1.5 tesla, a pair of overlapping stents at 1.5 tesla, a single stent at 3 tesla, and a pair of overlapping stents at 3 tesla, respectively) at whole-body averaged specific absorption rates (SAR) of 1.5 W/kg (for a single stent at 1.5 tesla) and 3 W/kg (for a single stent at 3 tesla and a pair of overlapping stents at 1.5 and 3 tesla) for 20 minutes (for a single stent at 1.5 tesla) and 15 minutes (for a single stent at 3 tesla and a pair of overlapping stents at 1.5 and 3 tesla) of MR scanning in a 1.5 tesla/64 MHz GE Healthcare MR scanner, a 1.5 tesla Magnetom Siemens Medical Solutions MR Scanner (to evaluate a pair of overlapping stents), and a 3 tesla Excite GE Healthcare MR scanner.

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the Zilver Self-Expanding Stent. Therefore, it may be necessary to optimize MR imaging parameters for the presence of this metallic implant.

    Heating in the MRI environment for stents with fractured struts is not known.

  • Embolization Coils - Stainless Steel

    Stainless steel coils are ferrous and are affected by the MR field. Benchtop testing on stainless steel embolization coil samples showed high degrees of torque and deflection. Therefore, stainless steel embolization coils have a designation of MR Unsafe and we would not recommend scanning a patient implanted with a stainless steel embolization coil.

  • Gianturco-Roehm Bird’s Nest® Vena Cava Filter

    Nonclinical testing has demonstrated that the Gianturco-Roehm Bird's Nest Vena Cava Filter is MR Conditional. According to the article “MR Imaging of the Bird's Nest Filter” by Alyssa T. Watanabe, “No complication or symptomatic filter displacement was encountered as a result of MR imaging performed at 1.5 T.” The article also states that the filters created significant local artifact distortion on MR images; however, diagnostic images of the pelvis, spine and brain may still be obtained.

    MRI diagnostic procedures should be postponed for six weeks following filter implantation to ensure device incorporation into the vessel wall.

  • Formula 418® Balloon Expandable Biliary Stent

    Nonclinical testing has demonstrated that the Formula biliary stent is MR Conditional. A patient with this stent may undergo MRI immediately after placement under the following conditions:

    • Static magnetic field of 3.0 tesla or less
    • Maximum spatial gradient magnetic field of 720 gauss/cm
    • Maximum MR system reported whole-body averaged specific absorption rate (SAR) of 3.0 W/kg for 15 minutes of scanning

    In nonclinical testing, the Formula biliary stent produced a temperature rise of less than 1°C at an MR-system-reported maximum whole-body averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of MR scanning in a 3 tesla system (Excite, Software G3.0-052B, GE Healthcare).

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the Formula biliary stent. Therefore, it may be necessary to optimize MR imaging parameters for the presence of this metallic implant. The effect of heating in the MRI environment for overlapping stents or stents with fractured struts is unknown.

  • Günther Tulip™ Vena Cava Filter

    Nonclinical testing has demonstrated that the Günther Tulip Vena Cava Filter is MR Conditional. It can be scanned safely immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient field of 525 gauss/cm or less
    • Maximum whole-body averaged specific absorption rate (SAR) of 3.0 W/kg for 15 minutes of scanning

    In nonclinical testing, the Günther Tulip Vena Cava Filter produced a temperature rise of less than 0.6°C at a maximum whole-body averaged specific absorption rate (SAR) of 3.0 W/kg for 15 minutes of MR scanning in a 3.0 tesla MR system using a transmit/receive body coil (Excite, Software G3.0-052B, GE Healthcare, Milwaukee, WI).

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the Günther Tulip Vena Cava Filter. Therefore, it may be necessary to optimize MR imaging parameters for the presence of this metallic implant.

  • Hilal Embolization Microcoils™

    Hilal Embolization Microcoils are manufactured from a platinum composition that has been shown to have no ferromagnetic response. Through nonclinical testing, embolization coils manufactured of this material have been shown to be MR Conditional. A patient with one of these implants can be safely scanned immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient field of 330 gauss/cm
    • Maximum MR system whole-body averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of scanning

    In this testing, platinum embolization coils produced a temperature rise of greater than 0.5°C at a maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI. The effect of heating for other conditions, multiple coils or overlapping coils is unknown.

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of platinum embolization coils.

  • Cook-Swartz Doppler Flow Probe

    WARNING–MR UNSAFE

    Do not expose patient to an MRI procedure while a Cook-Swartz Doppler Flow Probe is implanted. Substantial MRI-related heating of the Doppler flow probe may occur. Doppler flow probe must be removed prior to any MRI procedure.

  • MReye® Embolization Coils

    Through nonclinical testing, the MReye Embolization Coil has been shown to be MR Conditional. It can be scanned safely immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient of 330 gauss/cm
    • Maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI

    The MReye Embolization Coil will not migrate in this MRI environment. Nonclinical testing has not been performed to rule out the possibility of device migration at static magnetic field strengths higher than 3.0 tesla and a maximum spatial gradient higher than 330 gauss/cm.

    In this testing, the MReye Embolization Coil produced a temperature rise of greater than or equal to 0.6°C at a maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI. The effect of heating in the MRI environment for other conditions, multiple coils or overlapping coils is unknown.

    MRI image quality may be compromised if the area of interest is in the exact same area or relatively close to the position of the MReye Embolization Coil.

  • MReye Flipper® Detachable Embolization Coils

    Through nonclinical testing, the MReye Flipper Detachable Embolization Coil Delivery System has been shown to be MR Conditional. It can be scanned safely immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient of 330 gauss/cm
    • Maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI

    Embolization coils manufactured of Inconel will not migrate in this MRI environment. Nonclinical testing has not been performed to rule out the possibility of device migration at static magnetic field strengths higher than 3.0 tesla and a maximum spatial gradient higher than 330 gauss/cm.

    In this testing, the MReye Flipper Embolization Coil produced a temperature rise of greater than or equal to 0.6°C at a maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI. The effect of heating in the MRI environment for other conditions, multiple coils or overlapping coils is unknown.

    MRI image quality may be compromised if the area of interest is in the same area or relatively close to the position of the MReye Flipper Embolization Coil.

  • MReye® Needles for Magnetic Resonance Imaging

    Cook’s interventional MReye Needles are used for initial puncture and positioning control as well as biopsies and other interventional procedures in combination with Magnetic Resonance Imaging.

    • MReye products are created from a nonferromagnetic material that helps prevent disturbing artifacts.
    • MReye needles are recommended for use with MRI scanners not exceeding 1.5 tesla.

    To see different-sized images of the needle, try these scanner parameters:

    • To see an image of the needle at a ratio of 1:1, select a spin echo sequence (TR/TE 200 ms/15 ms).
    • To see an image of the needle enlarged approximately 2:1, select a gradient echo sequence (TR/TE/Flip angle 336 ms/15 ms/90°).
  • Nester® Embolization Coils

    Nester Embolization Coils are manufactured from a platinum composition that has been shown to have no ferromagnetic response. Through nonclinical testing, embolization coils manufactured of this material have been shown to be MR Conditional. A patient with one of these implants can be scanned safely immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient field of 330 gauss/cm
    • Maximum MR system whole-body averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of scanning

    In this testing, platinum embolization coils produced a temperature rise of greater than 0.5°C at a maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI. The effect of heating for other conditions, multiple coils or overlapping coils is unknown.

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of platinum embolization coils.

  • Tornado® Embolization Coils

    Tornado Embolization Coils are manufactured from a platinum composition that has been shown to have no ferromagnetic response. Through nonclinical testing, embolization coils manufactured of this material have been shown to be MR Conditional. A patient with one of these implants can be scanned safely immediately after placement under the following conditions:

    • Static magnetic field strengths of 3.0 tesla or less
    • Maximum spatial gradient field of 330 gauss/cm
    • Maximum MR system whole-body averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of scanning

    In this testing, platinum embolization coils produced a temperature rise of greater than 0.5°C at a maximum whole-body averaged specific absorption rate (SAR) of 2.0 W/kg for 20 minutes of MRI. The effect of heating for other conditions, multiple coils or overlapping coils is unknown.

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of platinum embolization coils.

  • Vital Port® System

    Nonclinical testing demonstrated that the Vital-Port Vascular Access Ports are MR Conditional. A patient with one of these implants can be safely scanned immediately after placement under the following conditions:

    • Static magnetic field of 3.0 tesla or less
    • Maximum spatial gradient magnetic field of 720 gauss/cm
    • Maximum MR system reported whole-body averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of scanning

    In nonclinical testing, the vascular access ports produced maximum temperature increases of less than or equal to 0.7°C at maximum MR system reported whole-body averaged specific absorption rate (SAR) of 3 W/kg for 15 minutes of MR scanning in a 3 tesla MR system using a transmit/receive body coil (Excite, software G3.0-052B, GE Healthcare, Milwaukee, WI).

    MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the vascular access ports. Therefore, it may be necessary to optimize MR imaging parameters to compensate for the presence of one of these implants.

  • Zenith TX2® TAA Endovascular Graft


      (excerpted from IFU)

    MRI Information
    Non-clinical testing has demonstrated that the Zenith TX2 TAA Endovascular Graft with Pro-Form is MR Conditional. It can be scanned safely under the following conditions:

    1.5 Tesla Systems:

    • Static magnetic field of 1.5 Tesla
    • Spatial gradient field of 450 Gauss/cm
    • Maximum whole-body-averaged specific absorption rate (SAR) of 2 W/kg for 15 minutes of scanning.

    In non-clinical testing, the Zenith TX2 TAA Endovascular Graft with Pro-Form produced a temperature rise of less than 1.4 °C at a maximum whole-body-averaged specific absorption rate (SAR) of 2.8 W/kg for 15 minutes of MR scanning in a 1.5 Tesla Magnetom, Siemens Medical Magnetom MR scanner. The maximum whole-body-averaged specific absorption rate (SAR) was 2.8 W/kg, which corresponds to a calorimetry measured value of 1.5 W/kg.

    3.0 Tesla Systems:

    • Static magnetic field of 3.0 Tesla
    • Spatial gradient field of 720 Gauss/cm
    • Maximum whole-body-averaged specific absorption rate (SAR) of 2 W/kg for 15 minutes of scanning

    In non-clinical testing, the Zenith TX2 TAA Endovascular Graft with Pro-Form produced a temperature rise of less than 1.9 °C at a maximum whole-body-averaged specific absorption rate (SAR) of 3.0 W/kg for 15 minutes of MR scanning in a 3.0 Tesla, Excite, GE Electric Healthcare MR scanner. The maximum whole-body-averaged specific absorption rate (SAR) was 3.0 W/kg, which corresponds to a calorimetry measured value of 2.8 W/kg.

    The image artifact extends throughout the anatomical region containing the device, obscuring the view of immediately adjacent anatomical structures within approximately 20 cm of the device, as well as the entire device and its lumen, when scanned in non-clinical testing using the sequence: Fast spin echo in a 3.0 Tesla, Excite, GE Electric Healthcare, with G3.0-052B software, MR system with body radiofrequency coil.

    For all scanners, the image artifact dissipates as the distance from the device to the area of interest increases. MR scans of the lower extremities may be obtained without image artifact. Image artifact may be present in scans of the abdominal, upper extremity, and head and neck region, depending on distance from the device to the area of interest.

    Clinical information is available on six patients who received MRI scans during the course of the clinical trial. There have been no reported adverse events or device problems in any of these patients as a result of having received an MRI. Additionally, there have been approximately 3,000 patients implanted with Zenith TAA Endovascular Grafts worldwide, in which there have been no reported adverse events or device problems as a result of MRI.

    Cook recommends that the patient register the MR conditions disclosed in this IFU with the MedicAlert Foundation. The MedicAlert Foundation can be contacted in the following manners:

    Mail: MedicAlert Foundation International
    2323 Colorado Avenue
    Turlock, CA 95382
    Phone: 888.633.4298 (toll free)
    209.668.3333 from outside the U.S.
    Fax: 209.669.2450
    Web: www.medicalert.org